icon
Specification

- Four channels and six channels cover most fluorescence;

Effectively reduces multi-color crosstalk and edge effects without fluorescence correction;

Innovative scanning method and time-resolved signal separation technology ;

Optical scanning detection system, high sensitivity and resolution, accurateidentification of negative and positive;

Unique zoned temperature control and edge temperature compensation technology ensures temperature uniformity.


Model

QP96C4

QP96C6

Sample capacity

96x0.1ml PCR tube/12x0.1ml 8-strip tube/96-well microplate

Reaction volume

10~50μl

Thermal cycling

Peltier(6 temperature control modules)

Maximum ramp rate

6℃/s

Temperature setting range

4~100℃

Temperature accuracy

≤ ±0.1℃

Temperature uniformity

≤±0.1℃

Gradient zone

12 columns

Gradient range

1~40°C

Excitation light source

4pcs LED

6pcs LED

Detection module

MPPC

Scanning principle

Time-resolved scanning technology

Excitation/detection range

415-685nm/455-745nm

Fluorescence channel

4 channels

6 channels

Detection sensitivity

415-685nm/455-745nm

Excitation/detection range

1 copy of the target sequence

System sensitivity

1.33-fold copy number differences in singleplex reactions

Dynamic range

10 orders of magnitude

Dye compatibility

F1: FAM/SYBR Greenl          F2: VIC/HEX/TET/JOE/TAMRA 

F3: ROX/Texas Red              F4: Cy5/LIZ       

F5: Quasar 705/AlexaFluor 680/Cy5.5 

F6:  ATTO425

Dimension

L500xW360xH380mm

Weight

25kg

Power supply

100-240V,50-60Hz